Experimental instrument
Test tube, Micropipettor, Vortex mixer, Low-speed centrifuge, 37℃ water bath/gas bath, Spectrophotometer (420nm)
Product introduction
This kit is based on the reaction of hydrogen peroxide catalyzed by peroxidase, it detects the enzymatic activity by measuring the diversification of the absorbency value at 420 nm.
Application
This kit can be used to measure the POD activity in animal serum (plasma) and other samples.
Operation steps
| Blank tube | Sample tube |
Reagent 1(mL) | 2.4 | 2.4 |
Reagent 2 application solution(mL) | 0.3 | 0.3 |
Reagent 3 application solution(mL) | 0.2 | 0.2 |
Double-distilled water(mL) | 0.1 |
|
Sample(mL) |
| 0.1 |
Incubate for accurately 30 min at 37℃. | ||
Reagent 4(mL) | 1.0 | 1.0 |
Mix fully and centrifuge for 10 min at 3500 rpm. Take the supernatant, then measure the OD values of each tube at 420 nm wavelength with 1cm diameter cuvette and set to zero with double-distilled water.