The principle of the ABTS method for determining the T-AOC is as follows. ABTS is oxidized to green ABTS+ by appropriate oxidant, which can be inhibited if there exist antioxidants. The T-AOC of the sample can be determined and calculated by measuring the absorbance of ABTS+ at 414 nm or 734 nm. Trolox is an analog of VE and has a similar antioxidant capacity to that of VE. Trolox is used as a reference for other antioxidant antioxidants. For example, the T-AOC of Trolox is 1, then the antioxidant capacity of the other substance with the same concentration is showed by the ratio of its antioxidant capacity to Trolox antioxidant capacity.
If the sample to be tested is water-soluble, dilute the standard and samples with PBS. If the sample to be tested is water-insoluble, dilute the standard and samples with 80% ethanol.
| 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 |
A | A | A | S1 | S9 | S17 | S25 | S33 | S41 | S49 | S57 | S65 | S73 |
B | B | B | S2 | S10 | S18 | S26 | S34 | S42 | S50 | S58 | S66 | S74 |
C | C | C | S3 | S11 | S19 | S27 | S35 | S43 | S51 | S59 | S67 | S75 |
D | D | D | S4 | S12 | S20 | S28 | S36 | S44 | S52 | S60 | S68 | S76 |
E | E | E | S5 | S13 | S21 | S29 | S37 | S45 | S53 | S61 | S69 | S77 |
F | F | F | S6 | S14 | S22 | S30 | S38 | S46 | S54 | S62 | S70 | S78 |
G | G | G | S7 | S15 | S23 | S31 | S39 | S47 | S55 | S63 | S71 | S79 |
H | H | H | S8 | S16 | S24 | S32 | S40 | S48 | S56 | S64 | S72 | S80 |
Note: A:blank wells, B-H:standard wells, S1-S80:sample wells.
1) Standard tube: Add 10 μL of standard with different concentration to the 96 well micro-plate.
Sample tube: Add 10 μL of sample to the 96 well micro-plate.
2) Add 20 μL of reagent 4 application solution to each well.
3) Add 170 μL of ABTS working solution to each well.
4) Mix fully and stand for 6 min at room temperature. Measure the OD values of each well at 414 nm with Microplate reader.
| Standard well | Sample well |
Standards with different concentrations (μL) | 10 |
|
Sample (µL) |
| 10 |
Reagent 4 application solution (μL) | 20 | 20 |
ABTS working solution (μL) | 170 | 170 |
Mix fully and stand for 6 min at room temperature. Measure the OD values of each well at 414 nm with Microplate reader. |
Detection range | 0.047-1.50 mmol/L | Average inter-assay CV | 4.1% |
Sensitivity | 0.047 mmol/L | Average intra-assay CV | 2.2% |
Average recovery rate | 101% |
상품문의가 없습니다.
추가비용을 고객에게 부담시지키 않습니다. (단, 고객 변심 또는 주문 반복으로 인한 경우의 반환비용은 고객님이 부담하셔야
합니다.)
::: 교환 및 반품이 가능한 경우:::
단, 상품을 개봉하여 상품가치가 상실된 경우에는 교환/반품이 불가능합니다.
:::교환 및 반품이 불가능한 경우:::
주문 취소 및 반품으로 환불을 요청하실 경우에는 E-mail(celgen-bio@celgen-bio.com)이나 고객만족센터 (042-824-9026)을
통해 요청하시면 친절하게 처리해 드리겠습니다.
주문 취소 후 반품 가능 여부를 확인한 다음 3일 이내에 결제 금액을 환불해 드리겠습니다.