[E-BC-K205] Low-density Lipoprotein Cholesterol (LDL-C) Colorimetric Assay Kit (Double Reagents) 요약정보 및 구매

96T

상품 선택옵션 0 개, 추가옵션 0 개

제조사 Elabscience
납기일 1-2 주
판매가격 210,000원
포인트 0점(구매금액 : 0%)
배송비결제 3,500원   (주문금액 10만원 이상일시 배송비 무료)

선택된 옵션

  • [E-BC-K205] Low-density Lipoprotein Cholesterol (LDL-C) Colorimetric Assay Kit (Double Reagents) (+0원)

상품 정보

상품 기본설명

96T

상품 상세설명

Application

This kit can be used for detection of low-density lipoprotein cholesterol (LDL-C) content in serum, plasma, cells, culture supernatant and tissue samples. 

 

Detection principle 

                                                                          

 

The coloured substance have a maximum absorption peak at 546 nm. Measure the OD value at 546 nm and the LDL-C content in the sample can be calculated.

 

Experimental instrument

 

Test tube, Micropipettor, Homogenizer, 37℃ water bath/gas bath, Microplate reader or Biochemical analyzer

 

Sample preparation

1. Serum (Plasma): Detect the sample directly. If the concentration is beyond the linear range, then dilute the sample with saline before detection.

2. Cell culture supernatant: Centrifuge at 1000 rpm for 10 min and then take the supernatant for test.

[Note]: It is generally recommended that the cell density should be more than 1×106/mL.

3. Tissue sample: Accurately weigh the tissue sample, add the homogenate medium at the ratio of Weight (g): Volume (mL) = 1: 9. Make the mechanical homogenization in ice water bath to prepare 10% homogenate. Centrifuge at 2500 rpm for 10 min, then take the supernatant for detection.

[Note]: (1) For high fat sample, the homogenate medium should be absolute alcohol.

(2) For non-high fat sample, the homogenate medium should be phosphate buffer (0.1 mol/L, pH 7.4) or normal saline.

4. Cell sample:

(1)  Cell collection: Take the prepared cell suspension and centrifuge at 1000 rpm for 10 min. Discard the supernatant and keep the cell sediment. Wash the sediment with iso-osmia buffer (0.1mol/L, pH7~7.4 phosphate buffer was recommended1~2 times, centrifuge at 1000 rpm for 10min, then discard the supernatant and keep the cell sediment.

(2)  Cell disruption: Add 0.2~0.3 mL of homogenate medium (0.1 mol/L, pH7~7.4 phosphate buffer or normal saline was recommended). Then treat the sample with sonication on ice (power: 300 W, 3~5 second/time, interval for 30 sec, repeat for 3~5 times) or grind with hand-operated. Take the prepared homogenate for detection without centrifugation. The cell can also be lysed with the cell lysate buffer (Triton X-100, 1~2%, lyse for 30~40 min), then take the prepared sample for detection without centrifugation. 

[Note]: It is generally recommended that the cell density should be more than 1×106/mL. The disrupted cell can be observed with microscope to check that whether the cell is broken completely.

 

Operation steps

1. Operate with 96 wells microplate.

 

Blank well

Standard well

Sample well

Distilled water (μL) 

2.5

 

 

Standard (μL)

 

2.5

 

Sample (μL)

 

 

2.5

Reagent 1 (μL)

180

180

180

Mix fully and incubate at 37 for 5 min. Measure the OD value (A1) at 546 nm with Microplate Reader. 

Reagent 2 (μL)

60

60

60

Mix fully and incubate at 37 for 5 min. Measure the OD value (A2) at 546 nm with Microplate Reader. 

 

2. Operate with Automatic biochemical analyzer.

a. Setting parameter

Main wavelength

546 nm

Reaction type

Terminal method

Reaction direction

Up reaction (+)

 

b. Operation steps

Sample/ Distilled water (μL)

2.5

Reagent 1 (μL)

180

Mix fully and incubate at 37 for 5 min. Measure the OD value (A1) at 546 nm with biochemical analyzer.

Reagent 2 (μL)

60

Mix fully and incubate at 37 for 5 min. Measure the OD value (A2) at 546 nm with biochemical analyzer.

 

Performance index

1. The absorbance of the blank tube is  0.050 (Optical path is 0.5 cm).

2. Sensitivity: The difference of absorbance value A is between 0.180~0.280 when testing 2.6 mmol/L samples.

3. Linear range: 0.02~12 mmol/L, r> 0.995.

4. Precision: the intra-assay CV ≤ 8%, the inter-assay CV≤ 10%.

5. Stability: This kit can be stored at 2~8℃ with shading light for 12 months and can be stable for 1 month since it is opened with the same storage condition.

 

Notes

1. This kit is for research use only.

2.  Instructions should be followed strictly, changes of operation may result in unreliable results.

3. The validity of kit is 12 months.

4. Do not use components from different batches of kit.

5. If the sample content is beyond the maximum limit, please dilute the sample with normal saline before detection, then multiply by the dilution ratio when calculate the result.

6. Protect the reagent from contamination of glucose, cholesterol, etc.

7. The amount of reagent and sample can be increased and decreased according to the requirement of automatic biochemical analyzer. 

상품문의

등록된 상품문의

상품문의가 없습니다.

반품/교환정보

셀젠바이오샵에서는 다음과 같은 기간 및 내용으로 상품에 대하여 교환, 반품, 환불을 보장하고 있으며, 상품의 반환에 의한

추가비용을 고객에게 부담시지키 않습니다. (, 고객 변심 또는 주문 반복으로 인한 경우의 반환비용은 고객님이 부담하셔야

합니다.)

 

::: 교환 및 반품이 가능한 경우:::

-상품을 공급 받으신 날로부터 7일 이내

  , 상품을 개봉하여 상품가치가 상실된 경우에는 교환/반품이 불가능합니다.

-공급받으신 상품 및 용역의 내용이 표시, 광고 내용과 다르거나 다르게 이행된 경우
-

:::교환 및 반품이 불가능한 경우:::

-하자가 없는 상품을 택배사로부터 인도 받으신 날로부터 20일 이후.
- 고객님의 책임 있는 사유로 상품등이 멸실 또는 훼손된 경우.
- 포장을 개봉하였거나 포장이 훼손되어 상품가치가 현저히 감소한 경우
- 시간의 경과에 의하여 재판매가 곤란할 정도로 상품등의 가치가 현저히 감소한 경우
- 주문 당시 해외 발주 상품의 경우
-

주문 취소 및 반품으로 환불을 요청하실 경우에는 E-mail(celgen-bio@celgen-bio.com)이나 고객만족센터 (042-824-9026)

통해 요청하시면 친절하게 처리해 드리겠습니다.

주문 취소 후 반품 가능 여부를 확인한 다음 3일 이내에 결제 금액을 환불해 드리겠습니다.
  • ㈜셀젠바이오테크, 대표자명: 김성빈, 사업자등록번호: 344-08-00018, 통신판매업종 제 2017-대전유성-0220 호
  • 대전광역시 대덕구 대화로69, 3층(대화동)
  • 대표전화: 042-824-9026 팩스: 042-824-9027
  • E-Mail: celgen-bio@celgen-bio.com
  • Copyright © 2015 Celgen-Bio.tech All rights reserved.