Application
The kit is used for the determination of Homocysteine (HCY) in serum samples.
Detection significance
The kit is used for auxiliary diagnosis of related diseases by determining the serum homocysteine concentration. Homocysteine is mainly used as a risk indicator of cardiovascular disease, especially coronary atherosclerosis and myocardial infarction. The increase in homocysteine concentration is proportional to the risk of disease and is an independent risk factor to induce cardiovascular disease.
Detection principle
Oxidized homocysteine (HCY) is reduced to free homocysteine by triethyl phosphine (TCEP), and the free homocysteine reacts with substrate to generate adenosine. The generated adenosine is immediately dehydrogenated into inosine and ammonia, and the ammonia is further react with NADH under the catalysis of glutamate dehydrogenase to convert NADH to NAD+. The decrease in absorbance at 340 nm caused by the decline of NADH is proportional to the concentration of homocysteine in the sample.
Applicable instruments
Automatic biochemical analyzer, incubator
Sample requirements
Collect the fasting serum by routine method. The sample is stable at 2-8℃ for 1 week and stable at -20℃ for several months. Do not use serum or plasma containing sodium fluoride. The Sample with hemolysis, turbidity, or severe blood lipid are not suitable for HCY detection. Try to avoid high protein diet before blood collection, which can lead to elevated HCY.
Operation steps
1. Detection with Biochemical analyzer
Setting parameter
Temperature | 37℃ | Method | Rate method |
Reaction direction | Down | Reaction time (Serum + R1) | 240 s |
Calibration method | Linear | Reaction time (Serum + R1+R2) | 120 s |
Sample volume | 13 μL | Dominant wavelength | 340 nm |
Reagent Ⅰ (R1) | 240 μL | Auxiliary wavelength | 405 nm |
Reagent Ⅱ (R2) | 65 μL | Detection time | 120 s |
Automatic biochemical analyzer has its own program parameter input language. Reagents matches the analyzer and carry out automatic measurement after the above basic parameters are modified.
2. Detection with Spectrophotometer
Operation table
| Sample tube | Blank tube | Standard tube |
Sample (μL) | 39 |
|
|
Standard Ⅰ (μL) |
| 39 |
|
Standard Ⅱ (μL) |
|
| 39 |
Reagent 1 (R1) (μL) | 720 | 720 | 720 |
Mix fully and incubate at 37℃ for 4 min. | |||
Reagent 2 (R2) (μL) | 195 | 195 | 195 |
Mix fully and incubate at 37℃ for 2 min. Set to zero with distilled water and measure the OD value at 340 nm with a 1 cm optical path cuvette. The OD value of 0 min and 2 min were recorded as A1 and A2, respectively. △A=A1-A2. Calculate ΔA/min = (A1-A2)/2 min. |
Reference range
0-15 μmol/L (This is for reference only.)
Performance parameters
1. A340 of blank ≥ 1.000.
2. ΔA/min of blank ≤ 0.0300 (340 nm, 1 cm optical path).
3. Sensitivity: The difference of absorbance value △A is less than 0.0100 when testing 10 μmol/L samples.
4. Linear range: 0-50 μmol/L, r2 ≥ 0.990.
5. The intra-assay CV ≤ 8 %, the inter-assay CV≤ 10 %.
6. The relative deviation is -15%~15%.
7. Stability: This kit can be store at 2-8℃ with shading light for 12 months. It can be stable for a month at 2-8℃ with shading light after opening.
Notes
1. This kit is for research use only.
2. Instructions should be followed strictly, changes of operation may result in unreliable results.
3. Do not use components from different batches of kit.
4. Do not mix Standard Ⅰ and Standard Ⅱ. Do not use components from different batches of kit.
5. Take the needed amount of reagents and keep the remaining reagent sealed in the refrigerator.
6. The sample needs to be diluted with normal saline before determination once the concentration is beyond the linear range. The result should be multiplied by the dilution factor.
7. Wear rubber gloves when using reagentⅡwhich contains sodium azide. It should be avoided to contact with skin and clothing. Wash immediately with plenty of water if contact it carelessly and seek for medical treatment if necessary. Other wastes should be treated according to relevant regulations.
상품문의가 없습니다.
추가비용을 고객에게 부담시지키 않습니다. (단, 고객 변심 또는 주문 반복으로 인한 경우의 반환비용은 고객님이 부담하셔야
합니다.)
::: 교환 및 반품이 가능한 경우:::
단, 상품을 개봉하여 상품가치가 상실된 경우에는 교환/반품이 불가능합니다.
:::교환 및 반품이 불가능한 경우:::
주문 취소 및 반품으로 환불을 요청하실 경우에는 E-mail(celgen-bio@celgen-bio.com)이나 고객만족센터 (042-824-9026)을
통해 요청하시면 친절하게 처리해 드리겠습니다.
주문 취소 후 반품 가능 여부를 확인한 다음 3일 이내에 결제 금액을 환불해 드리겠습니다.