Application
This kit can be used to measure direct bilirubin (D-BIL) content in serum or plasma.
Detection significance
High total bilirubin in serum (plasma) is common in hepatitis, extrahepatic biliary obstruction, hemolytic disease, neonatal jaundice, etc..
Detection principle
Direct bilirubin is oxidized under vanadonic acid to form biliverdin. The decrease of absorbance at 450nm is proportional to the content of direct bilirubin..
Experimental material
Microplate (96 wells), High-precision transferpettor, Microplate Reader (450 nm)
Operation steps (Microplate reader)
| Blank well | Sample well |
ddH2O (µL) | 7 |
|
Sample (µL) |
| 7 |
Reagent 1 (µL) | 200 | 200 |
Mix fully, react for 5min at room temperature and measure the OD value(A1) at 450 nm. | ||
Reagent 2 (µL) | 50 | 50 |
Mix fully, react for 5min at room temperature and measure the OD value(A2) at 450 nm. Calculate the △A=A1-A2. |
The reference range
Healthy human serum: 0-6.84 µmol/L (0-0.04 mg/dL) (This is for reference only.)
Limitations
1. Interferential substance: Hemoglobin≤0.2 g/L,Vitamin C≤30 mg/dL, Glycerin trilaurate≤2000 mg/dL, have no interference to the result.
2. direct bilirubin is just one of the indicators of the experiment. It is also necessary to make a comprehensive judgment according to the physical signs of the sample and other experimental indexs.
Product performance index
1. The OD value of blank is not more than 0.05..
2. Linear range : 1.0 - 410 µmol/L (a: R2≥0.995. b: 1.0-150 µmol/L, deviation from linearity is less than 20.0 µmol/L. c: 150-410 µmol/L, deviation from linearity is less than 10% ).
3. Precision: the CV% is less than 10%.
4. Recovery: 100±20%.
5. Sensitivity: When the concentration of sample is 50 µmol/L, the difference of absorbance (△A) is more than 0.04.
Notes
1. The working solution is corrosive, it should be avoided to contact with skin and clothing. Wash immediately with plenty of water if contact it carelessly.
2. Instructions should be followed strictly.
3. All the samples and waste material should be treated as infective material according to the relevant rules of biosafety.
4. Bilirubin can be easily decomposed by light and it should be avoid light when measure.
상품문의가 없습니다.
추가비용을 고객에게 부담시지키 않습니다. (단, 고객 변심 또는 주문 반복으로 인한 경우의 반환비용은 고객님이 부담하셔야
합니다.)
::: 교환 및 반품이 가능한 경우:::
단, 상품을 개봉하여 상품가치가 상실된 경우에는 교환/반품이 불가능합니다.
:::교환 및 반품이 불가능한 경우:::
주문 취소 및 반품으로 환불을 요청하실 경우에는 E-mail(celgen-bio@celgen-bio.com)이나 고객만족센터 (042-824-9026)을
통해 요청하시면 친절하게 처리해 드리겠습니다.
주문 취소 후 반품 가능 여부를 확인한 다음 3일 이내에 결제 금액을 환불해 드리겠습니다.