Nitrite and nitrate are important chemical components in almost all living organisms. In vivo, nitrite and nitrate come from the NO oxidation pathway without excessive uptake of nitrite and nitrate.
Nitrite can react with chromogenic agent producing light red azo-compound. The content of nitrite can be calculated indirectly by measuring the OD value at 550 nm.
The supernatant for chromogenic reaction should not contain sediment, otherwise it will affect the results.
1. Blank tube: Add a* mL of Double distilled water into a 2 mL EP tube.
Standard tube: Add a* mL of 100 μmol/L Sodium nitrite standard solution into a 2 mL EP tube.
Sample tube: Add a* mL of Sample into a 2 mL EP tube.
(a* = volume of sample = volume of standard = volume of double distilled water.)
(Reference sampling quantity: 0.2~0.4 mL for serum (plasma) and 0.1~0.2 mL for tissue or cells homogenate.)
2. Add 0.8 mL of Reagent 1 and oscillate fully with vortex mixer.
3. Add 0.4 mL of Reagent 2 and oscillate fully with vortex mixer.
4. Stand the tubes at room temperature for 10 min and centrifuge at 2000 g for 10 min. (The supernatant after centrifugation must be clarified, and if there is precipitate, it must be centrifuged again.)
5. Take 0.8 mL of supernatant into a 2 mL EP tube, add 0.4 mL of Chromogenic agent.
6. Mix fully and stand the tubes at room temperature for 15 min. Set to zero with double distilled water and measure the OD values of each tube at 550 nm with 0.5 cm diameter cuvette.
| Blank tube | Standard tube | Sample tube |
Double distilled water (mL) | a* |
|
|
100 μmol/L sodium nitrite standard solution (mL) |
| a* |
|
Sample (mL) |
|
| a* |
Reagent 1 (mL) | 0.8 | 0.8 | 0.8 |
Reagent 2 (mL) | 0.4 | 0.4 | 0.4 |
Mix fully and stand for 10 min at room temperature, centrifuge at 2000 g for 10 min, take the supernatant for detection. | |||
Supernatant (mL) | 0.8 | 0.8 | 0.8 |
Chromogenic agent (mL) | 0.4 | 0.4 | 0.4 |
Mix fully and stand the tubes at room temperature for 15 min. Set to zero with double distilled water and measure the OD values of each tube at 550 nm with 0.5 cm diameter cuvette. |
Detection range | 1.36-500 μmol/L | Average inter-assay CV | 3.3% |
Sensitivity | 1.36 μmol/L | Average intra-assay CV | 2.6% |
Average recovery rate | 97% |
상품문의가 없습니다.
추가비용을 고객에게 부담시지키 않습니다. (단, 고객 변심 또는 주문 반복으로 인한 경우의 반환비용은 고객님이 부담하셔야
합니다.)
::: 교환 및 반품이 가능한 경우:::
단, 상품을 개봉하여 상품가치가 상실된 경우에는 교환/반품이 불가능합니다.
:::교환 및 반품이 불가능한 경우:::
주문 취소 및 반품으로 환불을 요청하실 경우에는 E-mail(celgen-bio@celgen-bio.com)이나 고객만족센터 (042-824-9026)을
통해 요청하시면 친절하게 처리해 드리겠습니다.
주문 취소 후 반품 가능 여부를 확인한 다음 3일 이내에 결제 금액을 환불해 드리겠습니다.