Application
This kit can be used to measure the Albumin (ALB) content in serum, plasma samples.
Detection significance
Albumin is the most abundant plasma protein, with a molecular weight of 66.5 kDa. Albumin is synthesized in the liver at a rate of 9 to 12 grams per day and regulated by insulin, amino acid intake and low colloidal osmotic pressure. Changes of albumin content in urine and serum are predictors of diabetic nephropathy, cardiovascular disease, liver disease and sepsis.
Detection principle
Bromocresol green (BCG) is widely used as protein staining agent. BCG can combine the albumin in pH 4.0~4.2 to form an albumin-BCG complex. And the color changed from yellow to green. The depth of color is proportional to the concentration of albumin. The content of albumin in serum can be calculated indirectly by measuring the OD value at 630 nm.
Experimental instrument
Micropipette, Vortex mixer, Centrifuge, Microplate reader (630 nm)
Sample preparation
Serum (plasma) sample: detect directly.
Operation steps
It is recommended to take 2~3 samples which expected large difference to do pre-experiment before formal experiment. According to the results of pre-experiment and the linear range (0.08-15 g/L) of the kit, then dilute the samples.
1. Dilution of standard
Dilute 20 g/L standard solution with double distilled water to a serial concentration. The recommended dilution gradient is as follows: 15, 12, 8, 5, 3.5, 2, 1, 0 μg/mL.
2. Assay steps:
(1) Standard well: add 10 μL standard with different concentration into the wells.
Sample tube: add 10 μL of sample into the wells.
(2) Add 250 μL of the Reagent 1 working solution to each well.
(3) Stand for 10 min at room temperature.
(4) Measure the OD value of each well at 630 nm with Microplate reader.
Note: It can be refer to the following operating table.
| Standard well | Sample well |
Standard with different concentration (μL) | 10 |
|
Sample (μL) |
| 10 |
Reagent 1 working solution (μL) | 250 | 250 |
Stand for 10 min at room temperature. Measure the OD values of each tube at 630 nm with the Microplate reader. |
Technical parameters
1. The sensitivity of the kit is 0.08 g/L.
2. The intra-assay CV is 1.5 % and the inter-assay CV is 4.6 %.
3. The recovery of the kit is 95 %.
4. The linear range of the kit is 0.08 g/L-15 g/L.
Notes
1. The kit is for scientific research only.
2. Instructions should be followed strictly, changes of operation may result in unreliable results.
3. The validity of kit is 6 months.
4. Do not use components from different batches of kit.
상품문의가 없습니다.
추가비용을 고객에게 부담시지키 않습니다. (단, 고객 변심 또는 주문 반복으로 인한 경우의 반환비용은 고객님이 부담하셔야
합니다.)
::: 교환 및 반품이 가능한 경우:::
단, 상품을 개봉하여 상품가치가 상실된 경우에는 교환/반품이 불가능합니다.
:::교환 및 반품이 불가능한 경우:::
주문 취소 및 반품으로 환불을 요청하실 경우에는 E-mail(celgen-bio@celgen-bio.com)이나 고객만족센터 (042-824-9026)을
통해 요청하시면 친절하게 처리해 드리겠습니다.
주문 취소 후 반품 가능 여부를 확인한 다음 3일 이내에 결제 금액을 환불해 드리겠습니다.